Verifying Validation of the Anthrone Colorimetric Method for Rhamnolipids Quantification
Abstract
This paper aims to to validate by checking of anthrone colorimetric method to quantify rhamnolipids in microbial culture medium. Linearity, precision, accuracy, limits of detection and quantitation and robustness: This parameters were determined. It was demonstrated that the assay is linear and accurate in the range of 10-70 µg/mL of carbohydrates, coefficients lower than 3% variation for repeatability were obtained, and were found to have a detection limit equal to 1,63 mg / mL and a limit of quantification of 5,45 g / mL. It was found that the method is robust against variations in temperature and heating time as well as the amount of reagent added to the samples.Keywords: validation, anthrone, rhamnolipid, colorimetric method
Downloads
Published
2016-01-07
How to Cite
Rodríguez-Gámez, M. O., Vilasó-Cadre, J. E., Aguilera-Rodríguez, L. I., Pérez-Silva, D. R. M., & Ábalos-Rodríguez, D. A. (2016). Verifying Validation of the Anthrone Colorimetric Method for Rhamnolipids Quantification. Revista Cubana De Química, 25(3), 287–294. Retrieved from https://cubanaquimica.uo.edu.cu/index.php/cq/article/view/617
Issue
Section
Artículos
License
This journal provides immediate open access to its content, based on the principle that offering the public free access to research helps a greater global exchange of knowledge. Each author is responsible for the content of each of their articles.